7 Errors / Display messages
Operating Instructions DAC Universal
7.1 Display messages and what they mean
båÖäáëÜ
85
The pressure sensor is
defective.
Please contact your authorized dealer.
86
The safety ring was
activated when the lid
was closing.
Press the "Up" and "Right"-Keys simultaneously.
Make sure that the lid is placed properly in the lid
holder.
Restart the cycle.
If the error occurs repeatedly, please contact your
authorized dealer.
87
The safety ring is
defective.
Please contact your authorized dealer.
88
Vapor pressure was too
low during sterilization.
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC Universal.
If the error occurs repeatedly, please contact your
authorized dealer.
89
Vapor pressure was too
high during sterilization.
Please contact your authorized dealer.
Code
no.
Pictogram
Description
Corrective action