vPad-TI OPERATING MANUAL
5.12 Temperature Overshoot
It is inevitable that, during operation at
STC
, a care giver will be required to open a portal or a door
in the incubator and the incubator will lose temperature. The Standard prescribes how the incubator
should be able to recover from this problem in the
Temperature Overshoot
test in section 8,
clause 50.109. To simulate the problem described above, the incubator is operated at 32 °C until
STC
is reached (Figure 87). Once the temperature reaches
STC
(60 min+) at 30 °C the program
prompts the user to set the
Control Temperature
to 36 °C degrees, and the
Incubator
Temperature
is monitored (see Figure 88).
Figure 86
: Temperature Overshoot Instructions
Figure 87
: Initial Temperature Operation Prior to STC Complete
Running a Test/Chapter 5
#
Page 62