vPad-TI OPERATING MANUAL
To test the Skin Sensor, it must be placed in a tightly controlled temperature environment. vPad-TI
provides an optional Sensor Oven to perform this test. The Sensor is removed from the incubator
Baby Compartment and placed in the Oven. The Sensor remains plugged into the incubator
Control Panel and the incubator must be placed into the Skin Temperature Control Mode, with the
temperature of the Skin Sensor displayed. The adhesive disk which holds the sensor to the infant’s
skin must be removed and, if necessary, the sensor should be unrolled so the cable is straight. The
sensor should be positioned in the Oven as indicated in
Figure 43
with the sensor tip toward the
back of the thermal pad in the Oven, to provide good thermal contact.
The Oven is connected to the USB Host port of the tablet via a microUSB to microUSB cable, and
must be connected to its own 15-18 VDC power supply via the adapter input connection on the
back of the Oven. Press the Power ON/OFF switch on the front panel to activate the Oven.
When the Probe Test is initiated by touching the Probe Test icon, the app provides an instruction
screen on how to make the connection.
The
Skin Sensor Accuracy
Test can be performed at temperatures from 28 °C to 40 °C in
increments of 2 °C . The Oven is controlled by vPad-TI software when connected to the
USB
Host
port. Once the Sensor has been correctly placed in the Oven, as indicated above, click the
OK
button to initiate the Sensor Test. Alternately, the
Cancel
button can be clicked to abort the
Skin Sensor Accuracy
Test. During a test, the
Skin Sensor Accuracy
dialog box on the
screen can be reduced to the
Probe Test
icon by clicking on the
icon in the upper right
corner of the dialog box.
Figure 43
: Sensor Placed in Oven
Running a Test/Chapter 5
#
Page 39
Содержание vPad-TI
Страница 1: ...vPad TI Transport Incubator Tester Operating Manual MN 120 ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ...vPad TI OPERATING MANUAL 7 Calibration and Maintenance 73 ...
Страница 16: ...vPad TI OPERATING MANUAL Specifications Chapter 1 Page 4 ...
Страница 70: ...vPad TI OPERATING MANUAL Figure 79 Air Temperature Control Low Running a Test Chapter 5 Page 58 ...