vPad-TI OPERATING MANUAL
Once the mandatory information is entered, the OK button will be enabled. Touching the OK
button will begin the actual test.
As data is entered and accepted (by pressing OK) for the incubator inventory, a small database is
created which will allow an incubator to be selected from a list, in the future. To select an incubator
from the list, simply touch the Equipment ID field (or Serial Number field), and a list of pre-existing
entries will be displayed. Touch the entry which matches the incubator you wish to test, and all of
the fields will be populated from the database.
With the equipment information properly entered and the correct Autosequence selected, touch the
OK button to run the test. Details of the screens which will be displayed during the various
available tests will be covered in the next chapter.
Figure 26
: Equipment Information Page
Figure 27
Operation/Chapter 4
#
Page 32
Содержание vPad-TI
Страница 1: ...vPad TI Transport Incubator Tester Operating Manual MN 120 ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ...vPad TI OPERATING MANUAL 7 Calibration and Maintenance 73 ...
Страница 16: ...vPad TI OPERATING MANUAL Specifications Chapter 1 Page 4 ...
Страница 70: ...vPad TI OPERATING MANUAL Figure 79 Air Temperature Control Low Running a Test Chapter 5 Page 58 ...