vPad-TI OPERATING MANUAL
The Standard requires that the sound level of the alarm is at least 65 dBA at the prescribed position.
When the alarm sound measurement is complete, the result will be displayed as in Figure 97.
Figure 96
: Alarm Sound Level Instructions - Outside
Figure 97
: Alarm Sound level - Outside
Running a Test/Chapter 5
#
Page 67