vPad-TI OPERATING MANUAL
Abbreviations, Definitions and Symbols
The following abbreviations, terms and acronyms are used throughout this manual:
ACTRE
Air Control Temperature Regulation Error
AIT
Average Incubator Temperature
ATX
Average Temperature of Temperature Sensor X, where X = 2 - 5
AV
Air Velocity
BSL
Background Sound Level
°
C
degrees Celsius (centigrade)
°
F
degrees Fahrenheit
Autosequence A series of measurements or test operations that are run
automatically in a predefined order, with or without user involvement.
DUT
Device Under Test
GUI
Graphical User Interface
IT
Incubator Temperature
IIT
Independent Incubator Temperature
NSL
Noise Sound Level
STC
Steady Temperature Condition
STS
Skin Temperature Sensor
TC
Temperature Control
TEMP
Temperature
WT
Warm Up Time
CAUTION: Read the instructions for use before operating this device.
Definitions:
Some of the following definitions are paraphrased from the relevant sections and
clauses of IEC 60601-2-20, Edition 2
Air Control
Temperature
Regulation Error
Clause 201.12.1.106: an incubator operating in AIR
TEMPERATURE CONTROL shall have an AVERAGE
TEMPERATURE that does not differ from the CONTROL
TEMPERATURE of 36°C by more than ±1.5°C at an ambient
temperature between 20°C and 25°C; or, ±2°C at an ambient
temperature between 10°C and 20°C
Average Transport
Incubator
Temperature
Clause 201.3.203: the average of the maximum and minimum
INCUBATOR TEMPERATURES achieved during STEADY
TEMPERATURE CONDITION
Содержание vPad-TI
Страница 1: ...vPad TI Transport Incubator Tester Operating Manual MN 120 ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ...vPad TI OPERATING MANUAL 7 Calibration and Maintenance 73 ...
Страница 16: ...vPad TI OPERATING MANUAL Specifications Chapter 1 Page 4 ...
Страница 70: ...vPad TI OPERATING MANUAL Figure 79 Air Temperature Control Low Running a Test Chapter 5 Page 58 ...