vPad-TI OPERATING MANUAL
If the smaller
Temperature
Puck
seems too close, it can be rotated 180 degrees, in which
case, add 3.5" to the markings on the long bar. In the example above, the 9.5 inch marking (6"
+ 3.5" = 9.5") would be used.
Temperature Pucks for each of the other three quadrants are set up in the same manner. The
fully assembled system is shown in Figure 9
Figure 9
.
The vPad-TI Transport Incubator test system is now configured mechanically, according to the
Standard, to take temperature and other measurements in the center of the incubator and in the
center of each of the four quadrants of the mattress. It is now necessary to make the electrical
connections for each of the sensors.
As previously described, the large round ‘Puck’ represents a thermal mass imitating the infant
body in a convection warming environment. This ‘Puck’ is made of aluminum with a black
heat absorbing coating, with a temperature sensor embedded in the aluminum block. In this
configuration, the temperature measured by the ‘Puck’ is specifically configured to test
radiant
warmers
. For testing
incubators
, a Temperature Probe is inserted into the connector at the
center of each Puck. This Temperature Probe bypasses the thermistor in the Puck and elevates
the thermistor in the Probe to a position approximately 10 cm above the mattress, in
accordance with the Standard.
Figure 8
: Positioning Frame Assembly
Overview/Chapter 2
#
Page 10
Содержание vPad-TI
Страница 1: ...vPad TI Transport Incubator Tester Operating Manual MN 120 ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ...vPad TI OPERATING MANUAL 7 Calibration and Maintenance 73 ...
Страница 16: ...vPad TI OPERATING MANUAL Specifications Chapter 1 Page 4 ...
Страница 70: ...vPad TI OPERATING MANUAL Figure 79 Air Temperature Control Low Running a Test Chapter 5 Page 58 ...