vPad-TI OPERATING MANUAL
After
STC
is reached, it is possible to enter the Humidity and Temperature values indicated by the
incubator. The
Average Relative Humidity
value, as measured by the
Humidity Sensor
, is
displayed on the screen. If the incubator has its own indicator of humidity, the
Humidity Check
button can be clicked, enabling the input screen of Figure 75. The relative humidity value displayed
by the incubator is input and will be compared to the relative humidity as measured by the
Humidity Sensor
. The value indicated by the incubator must be within 15% of the value
measured by vPad-TI’s Humidity Sensor.
Figure 75
: Relative Humidity Input
Figure 76
: Independent Temperature Value Input
Similarly, the incubator is required to have a temperature sensor which is independent of the control
system of the incubator. This sensor is typically a non-mercury thermometer mounted within the
baby compartment of the incubator. The requirements of the Standard, clause 201.12.1.105, are
such that the independent temperature indicator must not differ from the Incubator Temperature by
more than 1
o
C. The indicated value of the independent temperature indicator is input into the
appropriate field in Figure 76.
Figure 77
Running a Test/Chapter 5
#
Page 56
Содержание vPad-TI
Страница 1: ...vPad TI Transport Incubator Tester Operating Manual MN 120 ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ...vPad TI OPERATING MANUAL 7 Calibration and Maintenance 73 ...
Страница 16: ...vPad TI OPERATING MANUAL Specifications Chapter 1 Page 4 ...
Страница 70: ...vPad TI OPERATING MANUAL Figure 79 Air Temperature Control Low Running a Test Chapter 5 Page 58 ...