vPad-TI OPERATING MANUAL
5.7 Air Temperature Control Regulation Error(ACTRE) at 36
°C
Working as an air controlled Transport Incubator, the average Transport Incubator Temperature
must not differ from the control temperature by more than ± 2°C at an ambient temperature of
between 10°C and 20°C and by no more that ±1.5°C at an ambient temperature of between 20°C
and 30°C.
Once the autosequence is selected and run, the system will display a Test Checklist
Set up the test environment and on pressing OK, the tablet will initiate communication with the
DAT inside the incubator. The tablet will then display the following set up parameters:
Once configured, pressing Continue will start the test by measuring the internal temperatures of the
incubator until STC is established
Figure 67
: ACTRE Checklist
Figure 68
: ACTRE Set Up Instructions
Running a Test/Chapter 5
#