vPad-TI OPERATING MANUAL
If the option to
Save Picture to JPG Afer Test
has been enabled in the
System Settings
, the
Result
will include a filename of the JPG image of the temperature versus time graph. This
filename will be included in the test report for reference.
5.6 Cold Ambient Temperature at 36
°C
The standard states that following changes in ambient temperature, a the Transport Incubator
Temperature must not differ from the Control Temperature by more than 3°C. To be able to verify
this, the ambient temperature must be in the range 21°C to 25°C and the incubator is heated up to
36°C. Once STC has been reached, the external power source is disconnected and the incubator is
moved to an environment where the ambient temperature is in the range -5°C (±2°C )
and
the
wind
velocity
is
not more than 1 m/s. After 15 minutes, the incubator is return to the original
environment (ambient temp 20°C to 25°C ), external power source reconnected and the incubator
run for a further 30 minutes. The Incubator Temperature must not vary by more than 3°C
throughout this test.
When initiated from an autosequence, a Test Checklist will be displayed on the screen:.
Figure 58
: Continue Warm Up Test
Running a Test/Chapter 5
#
Page 47
Содержание vPad-TI
Страница 1: ...vPad TI Transport Incubator Tester Operating Manual MN 120 ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ...vPad TI OPERATING MANUAL 7 Calibration and Maintenance 73 ...
Страница 16: ...vPad TI OPERATING MANUAL Specifications Chapter 1 Page 4 ...
Страница 70: ...vPad TI OPERATING MANUAL Figure 79 Air Temperature Control Low Running a Test Chapter 5 Page 58 ...