vPad-TI OPERATING MANUAL
In addition, the
Incubator (air) Temperature
and the
Mattress Temperature
will be
displayed on a minute to minute basis. The
Average Incubator Temperature
and the
Average Relative Humidity
will not be displayed until
STC
is reached. During the time before
STC
is true, the buttons labeled
Humidity Check
and
Temperature Check
are also disabled
(grayed out). After
STC
is true, these buttons will be enabled for input.
During the test, all of the data points collected, including those for the temperatures at the other 4
locations, are saved to a Raw Data file. All five temperatures for locations Ta through Te are
displayed along the bottom of the graph for information purposes. Any event which occurs during
the test is also saved as a comment. Typical comments would be: Minute Mode, Air Temperature
Control (Mattress Horizontal) at 36 °C Started, Waiting for STC, Humidity Check Pass, etc. The
Raw Data file can be exported and used to recreate the test graphics.
Figure 73
Figure 74
Running a Test/Chapter 5
#
Page 55