vPad-TI OPERATING MANUAL
vPad-TI monitors the
Average Incubator Temperature (IT)
and checks to see that the
Incubator Temperature
does not differ from
IT
by more than 1.5
o
C. vPad-TI will display the
length of time
STC
has been true using dashed green lines on the display. If this
STC
measurement is true within 15 minutes of first reaching 36 °C, the incubator passes the test
requirements. The
Results
will be shown in the dialog box as shown in Figures 91.
When the test is complete, the display will show the graphical result, from the start of the test. This
graph will be saved as a JPG image if the setting
Save Picture to JPG After Test
is checked. The
filename for the image will also be referenced in the test report.
Figure 90
: Higher Temperature STC Complete, Countdown in Progress
Figure 91
: Temperature Overshoot Test Results
Running a Test/Chapter 5
#
Page 64