Engström Carestation
8-18
08/07 1505-1018-000
8.6.6 Neon tools
The Neon Toolbox group contains lists of counts, flows, and zero offsets
related to the Neonatal Flow Sensor (NFS).
VCB flows
VMB flows
Item
Function
Unit
Insp Flow
Displays the Inspiratory Flow counts
reported by the Neonatal Flow Sensor
Counts
Exp Flow
Displays the Expiratory Flow counts
reported by the Neonatal Flow Sensor.
Counts
Net Flow
Net Flow is calculated from the counts
reported by the Neonatal Flow Sensor and
the current zero point. Inspiratory flow is
displayed as positive net flow and
expiratory flow is displayed as negative net
flow.
mL/min
Item
Function
Unit
Insp Flow
Displays the Inspiratory Flow counts
reported by the Neonatal Flow Sensor
Counts
Exp Flow
Displays the Expiratory Flow counts
reported by the Neonatal Flow Sensor.
Counts
Net Flow
Net Flow is calculated from the counts
reported by the Neonatal Flow Sensor and
the current zero point. Inspiratory flow is
displayed as positive net flow and
expiratory flow is displayed as negative net
flow.
mL/min