7 Troubleshooting
1505-1018-000 08/07
7-19
-none-
Neo Exh Flow Sensor
Comparison
High
(Exh TVexp - Neo TVexp) > (0.3* Neo
TVexp) or 30 mL whichever is greater
for 6 consecutive breaths.
• Calibrate neo flow sensor.
• Check neo flow sensor and cable connections.
• Calibrate expiratory flow sensor.
• Replace neo flow sensor.
• Replace neo flow sensor cable.
• Replace expiratory flow sensor.
• Replace VMB.
• Repace motherboard.
• Replace VCB.
-none-
Therapy Power Off
High
Power switch is set to Off.
Non-silenceable.
• Normal operation if On/Standby switch is moved to standby while in therapy.
• Verify vent is in standby state prior to turning power On/Standby switch to standby.
• Verify display unit encoder accepts other inputs reliably. If not, replace encoder.
• Check/replace harness to On/Standby switch.
• Check/replace On/Standby switch.
• Check/replace PMB.
Air supply
pressure high
Air Supply Pres High
Low
Air supply pressure > 95 psig for more
than 0.5 seconds
May result in FiO2 low alarm.
• Check air supply.
• Disconnect sensor and verify output is zero gauge pressure.
• Interchange harness connections on both supply pressure sensors.
• If alarm follows the sensor, replace supply pressure sensor.
• If not, check/replace VMB.
Air supply
pressure low
Flow Wave:
O2 only
Air Supply Pres Low
Medium
Air supply pressure < 24.3 psig for
more than 0.5 seconds
Unit shuts down Air supply and
delivers 100% Oxygen
• Check air supply.
• Disconnect sensor and verify output is zero gauge pressure.
• Interchange harness connections on both supply pressure sensors.
• If alarm follows the sensor, check sensor connector terminals. Replace supply pressure sensor if necessary.
• If not, check/replace harness.
• If not harness, check/replace VMB.
Alarm message
Alarm ID
Priority
Alarm Condition
Special Behavior/Comments
• Action/Troubleshooting
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...