8 Software Download and Diagnostics
1505-1018-000 08/07
8-23
8.7.1 VMB Channel
Configurations
The VMB Channel Configurations group includes items that are measured
from the VMB. If the item is within the acceptable range, “OK” is displayed
next to the value. If the item is out of the acceptable range, “FAIL” is displayed
instead.
Item
Description
Format
Units
Range
Expiratory Press
Expiratory Pressure sensor
measurement
XXX.XX
cmH2O
-26.8 to 131.8
Barometric Press
Barometric pressure sensor
measurement
XXX.XX
cmH2O
643.7 to 1182.0
O2 Supp Press
O2 supply pressure sensor
measurement
XXX.XX
psig
-9.1 to 127.8
Air Supp Press
Air supply pressure sensor
measurement
XXX.X
psig
-9.1 to 127.8
Reference
1.22V Reference
XXXXX
mV
1104 to 1338
12.5V Supply
12.5V Power supply to the VMB
XXXXX
mV
10900 to 13320
6.1V Expiratory
6.1V power supply to the expiratory
flow sensor
XXXXX
mV
5770 to 6370
5V O2 Supply
5V power supply to the O2 sensor
XXXXX
mV
4720 to 5380
5V O2 Press
5V power supply to the O2 pressure
sensor
XXXXX
mV
4680 to 5340
5V Air Press
5V power supply to the Air pressure
sensor
XXXXX
mV
4680 to 5340
5.5V Analog
5.5V analog supply
XXXXX
mV
5140 to 5880
-5.5V Analog
-5.5V analog supply
XXXXX
mV
-6220 to -4440
Ground
Ground
XXXXX
mV
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...