7 Troubleshooting
1505-1018-000 08/07
7-47
Low Pressure Leak Test
Complete
Vent engine low pressure leak test in progress.
None
Low Pressure Leak Test Fail
Vent engine low pressure leak test failed
because leak was too high.
Refer to Section Section 5.2.5
Measure Resis Chk
Breathing circuit resistance measurement in
progress.
None
Measure Resis Chk
Complete
Breathing circuit resistance measurement
completed successfully.
None
Measure Resis Chk Fail
Breathing circuit resistance measurement
failed.
See previous message in status log for cause of
failure.
Measured Resistance:
Value
Resistance measured during Checkout.
None; for information only.
Measuring P0.1
P0.1 procedure in progress
None
Minimum Flow not
achieved: Value
Could not achieve minimum flow required for
step in Checkout.
Value = maximum flow that was achieved
(mLpm)
1. Verify breathing circuit is connected and not
leaking.
2. Verify low pressure vent engine is not leaking.
3. Verify exhalation valve is properly connected
and all seals are in place.
4. Verify gas flow from flow control valves. Recal
if necessary.
Neo Flow Sensor Cal
Neonatal flow sensor calibration in progress.
None
Neo Flow Sensor Cal Fail
Neonatal flow sensor calibration failed.
1. Verify neonatal flow sensor is properly
connected and properly occluded.
Recalibrate the sensor.
2. Clean or replace neonatal flow sensor.
Neo Flow Sensor Cal Pass
Neonatal flow sensor calibration passed.
None
Neonatal Check Failed
(20 Lpm)
System was unable to achieve 20 Lpm of flow
during the Neonatal checkout.
1. Verify breathing circuit is connected and not
leaking.
2. Verify low pressure vent engine is not leaking.
3. Verify exhalation valve is properly connected
and all seals are in place.
4. Verify gas flow from control valves. Recal if
necessary.
Neonatal Check Failed
(Downstream)
The expiratory flow measured by the neonatal
flow sensor is more than 25% different from the
flow measured by the total flow sensor.
Refer to Section Section 7.1.11
Message in
“
Calibration running”
field
Reason
Troubleshooting
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...