Engström Carestation
7-36
08/07 1505-1018-000
Failed State
message:
System failure.
Service required
VMB to DU Com Error
Failed
Proper communications from the VMB
to the DU is not successfully
completed within a 10 second window.
Note
: Timing does not start for 90
seconds after power up.
Run Backup Ventilation if in
Therapy.
• Cycle power.
• Check error log for underlying causes such as voltage range errors.
• Using service card, check DU communication to VCB.
• If multiple subsystem communication failures:
-Check/replace DU cable.
-Verify DU system interface PCB is fully inserted into DU CPU PCB.
-Check/replace DU system interface PCB (connector panel assembly).
-Check/replace DU CPU board.
• If not multiple subsystem communication failures:
-Check/replace DU cable
-Verify VMB is fully inserted into mother board.
-Replace VMB.
-Check/replace motherboard.
-Check/replace DU system interface PCB (connector panel assembly).
-Check/replace DU CPU board.
Failed State
message:
System failure.
Service required
VMB to VCB Com Error
Failed
Proper communication of the high
speed sensor data from the VMB to the
VCB is not successfully completed
within a 50 ms window.
• Cycle power.
• Check error log for underlying causes such as voltage range errors.
• Check DU/VMB and DU/VCB communication using Service Card.
• Verify VMB and VCB are fully inserted into motherboard.
• Check/replace VMB.
• Check/replace VCB.
• Check/replace motherboard.
Alarm message
Alarm ID
Priority
Alarm Condition
Special Behavior/Comments
• Action/Troubleshooting
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...