Engström Carestation
7-30
08/07 1505-1018-000
Patient
connected?
Standby Patient
Detection
Suction patient
detection
High
Paw insp or Paw exp > 3 cmH2O when
bias flow set to 3 l/min air or O2 (if Air
not available) and exhalation valve set
to 6 cmH2O for 3 seconds.
Paw exp> 3 cmH2O for approx.
1 second
Or
Paw exp< -0.5 cm H2O for approx.
1 second
Or
Or Exp Flow< -1 l/min for 1 second
Non-silencable
Audible in Standby and
Monitoring states.
When placed in Standby state
either by the user or after
power up, the ventilator will
maintain a bias flow so as to
detect if a patient has been
attached to the circuit. This
functionality alerts for an
inadvertent attachment of the
patient without the user
actuating ventilation. The
Standby menu shall be
displayed with the highlight on
Start Ventilation when alarm is
initially activated.
• Was patient connected during standby state? If yes, this is a user error. If no:
• Remove all tube connections from vent. If alarm persists, check exhalation pressure port and exhalation
valve housing.
• Calibrate Flow control valves.
• Calibrate airway pressure transducers.
• Using the Service App and occluded circuit, set 3 l/min flow, and set 6 cmH20 on exh. Valve. Ensure both
Insp and Exp pressure sensors read 6 ±3 cmH20. Open circuit and verify pressure is < 2 cmH20.
• Check/replace zero valves.
• Replace appropriate pressure sensor board.
• Replace appropriate flow control valve.
Paux high
Paux High
Low
Paux > 100 cmH2O
De-energize Paux purge valve
(if purge valve is energized)
until alarm condition clears.
• Verify Paux port (and any tubing that is connected to Paux port) is not occluded when purge valve is energized
(purge is turned ON).
• Zero Paux pressure sensor.
• Use the Service App to calibrate the Paux pressure sensor (Section 5.2.8).
• Interchange Paux and Pexp connections. If Paux high alarm stops, replace Paux sensor.
• If Paux High alarm continues, replace VCB.
Paux sensor out
of range
Paux Sensor out of
range
Low
Paux sensor data out of range.
Range: 10-4095 mV
• Check/replace harness.
• Interchange Paux and Pexp connections. If Paux alarm stops, replace Paux sensor.
• If Paux alarm continues, replace VCB
Paux sensor out
of range
Paux Zeroing Error
Low
Range for zero pressure: 631-96
8
mV
(counts)
If failed, use previous known
good offset value for Paux.
• Verify Paux port (and any tubing that is connected to Paux port) is not occluded when purge valve is energized
(Purge is turned ON).
• Zero Paux pressure sensor.
• Check/replace harness.
• Interchange Paux and Pexp connections. If Paux alarm stops, replace Paux sensor.
• If Paux alarm continues, replace VCB
Alarm message
Alarm ID
Priority
Alarm Condition
Special Behavior/Comments
• Action/Troubleshooting
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...