Notes
6-4
08/07 1505-1018-000
6.2 Engström Carestation Planned Maintenance
6.2.1 Every twelve
(12) months
Perform the following steps every 12 months.
For details, refer to the sections listed.
#
1. If required, adjust the Display Unit arm joints (
Section 9.6
).
#
2. If required, adjust patient monitoring module rack arm.
#
3. Visually inspect Vent Engine fan filter. Clean or replace as needed.
#
4. Visually inspect Display Unit fan filter. Clean or replace as needed.
#
5. Visually inspect exhalation valve assembly. Clean or replace as needed.
#
6. Visually inspect gas inlet filters (O
2
and Air). Replace as needed.
CAUTION
Any visual signs of moisture or particulate matter indicates that the facility
may have a problem with their Medical gas delivery system. Return the
replaced filter to the appropriate authority in the facility so that they are aware
of the problem and can correct as needed. If there is any indication of
moisture in the filter, the main manifold must be inspected to ensure there are
no visual signs of moisture entering into the ventilator engine.
#
7. Complete the battery capacity test (
Section 6.5
).
#
8. Electrical safety tests (
Section 3.5
).
#
9. Complete all Service level tests and calibrations (
Section 5.2
).
#
10. Complete all Super-User calibrations (
Section 5.1
).
#
11. Complete the Checkout procedure (
Section 3.2
).
6.2.2 Every forty-eight
(48) months
In addition to the 12-month requirements, replace the following parts every
48 months (4 years).
#
1. Replace the system batteries* (Stock Number 1009-5682-000).
Note
: Refer to the “
Battery capacity test
” in
Section 6.5
.
#
2. Check regulator output (
Section 5.2.4
).
6.2.3 Every sixty (60)
months
In addition to the 12-month requirements, replace the following parts every
60 months (5 years).
#
1. Replace Lithium battery (1009-5800-000) on display CPU board.
Refer to
Section 9.5.2
, item 5.
Serial Number:
Date: (YY/MM/DD)
/
/
Hospital:
Performed by:
#
12 months
#
24 month
#
48 month
#
______________________
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...