9 Repair Procedures
1505-1018-000 08/07
9-11
9.3.7 To replace
Vent Engine
assembly,
complete
1. Remove the chassis from the housing. Refer to Section 9.3.1.
2. Remove the Vent Engine from the chassis. Refer to Section 9.3.2
3. Replace the Vent Engine with new Vent Engine assembly and connect harness to the
VCB and VMB.
N
N
N
Nooo
otttteee
e:::: Do not install Vent Engine assembly into chassis until all tests and calibration in
Step 4 have been completed.
4. Perform Service level tests and calibration. Refer to Section 5.2.
5. Install Vent Engine assembly into chassis. Refer to Section 9.3.2.
6. Install chassis into housing. Refer to Section 9.3.1.
7. Perform Super-User calibrations. Refer to Section 5.1.1.
8. Perform Automated Checkout. Refer to Section 3.2.
9.3.8 To replace
Vent Engine,
sub-assembly
1. Remove the chassis from the housing per Section 9.3.1.
2. Remove the Vent Engine assembly from the chassis. Refer to Section 9.3.2.
3. Remove the following components from the Vent Engine assembly:
a. Harness (vent engine to VCB/VMB) - (1505-5706-000).
b. Inspiratory Manifold Panel Assembly with safety valve (1505-8502-000).
c. Outlet Manifold Panel Assembly (including all mounted components) -
(1505-8505-000).
d. O2 Pipeline Adapter (from inlet manifold) - (1505-3206-000).
e. AIR Pipeline adapter (from inlet manifold) - (1505-3207-000).
4. Clean and install the above removed parts on the new Vent Engine sub assembly.
Refer to Section 9.3.4.
5. Connect the new Vent Engine assembly harness to the VCB and VMB.
N
N
N
Nooo
otttteee
e:::: Do not install Vent Engine assembly into chassis until all tests and calibration in
Step 6 have been completed.
6. Perform Service level tests and calibration. Refer to Section 5.2.
7. Install Vent Engine Assembly into the chassis. Refer to Section 9.3.2
8. Install chassis into housing. Refer to Section 9.3.1.
9. Perform Super-User calibrations. Refer to Section 5.1.1.
10.Perform Automated Checkout. Refer to Section 3.2.
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...