7 Troubleshooting
1505-1018-000 08/07
7-51
Paracube Chk Fail (50/50
Air)
Paracube (O2 sensor) check failed to reach 15
Lpm air flow for 50/50 cal.
1. Verify breathing circuit is connected and not
leaking.
2. Verify low pressure vent engine is not leaking.
3. Verify exhalation valve is properly connected
and all seals are in place.
4. Verify gas flow from flow control valves. Recal
if necessary.
Paracube Chk Fail (Air)
Paracube (O2 sensor) check failed 21% cal.
Paracube never completed 21% calibration.
Check Paracube.
Paracube Chk Fail (O2)
Paracube (O2 sensor) check failed to reach 30
Lpm O2 flow for O2 cal.
1. Verify breathing circuit is connected and not
leaking.
2. Verify low pressure vent engine is not leaking.
3. Verify exhalation valve is properly connected
and all seals are in place.
4. Verify gas flow from flow control valves. Recal
if necessary.
Paracube Fail
Paracube (O2 sensor) check failed to reach 30
Lpm air flow.
1. Verify breathing circuit is connected and not
leaking.
2. Verify low pressure vent engine is not leaking.
3. Verify exhalation valve is properly connected
and all seals are in place.
4. Verify gas flow from flow control valves. Recal
if necessary.
Paracube Sensor Fail (50/
50 O2)
Paracube (O2 sensor) check failed to reach 15
Lpm O2 flow for 50/50 cal.
1. Verify breathing circuit is connected and not
leaking.
2. Verify low pressure vent engine is not leaking.
3. Verify exhalation valve is properly connected
and all seals are in place.
4. Verify gas flow from flow control valves. Recal
if necessary.
Paux Zeroing...
Auxiliary pressure sensor is being zeroed at start
of Span Cal.
None
Paw Xducer Check
Airway pressure sensor check in progress.
None
Paw Xducer Check
Complete
Airway pressure sensor check passed.
None
Paw Xducer Check Fail
Airway pressure sensor check failed.
See previous message in status log for cause of
failure.
Paw Xducer Check Fail
(PressureZeroCal)
Pressure sensors were not successfully zeroed.
Refer to Section 7.1.1 and Section 5.2.8.
Paw Xducer OK
Airway pressure sensor check passed.
None
Message in
“
Calibration running”
field
Reason
Troubleshooting
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...