Engström Carestation
9-30
08/07 1505-1018-000
9.6.2 Adjust lower
pivot
1. Position the DU as shown:
• Arm forward
• Upper and lower pivots
in line horizontally
• DU vertical
2. Release the DU.
• Verify that the DU does not sag
from its own weight.
3. Push down slightly at the top of the
DU.
• The DU should remain in place.
4. Verify that the DU can be raised and lowered without excessive force.
5. Adjust the lower pivot as necessary.
9.6.3 Adjust arm
bearing
1. Position the DU as shown above:
• Arm forward
• Upper and lower pivots
in line horizontally
• DU vertical
2. Move the DU side to side.
• Verify that the DU can be moved
without excessive force and that it
remains in place with slight
sideways pressure.
3. Adjust the bearing as necessary.
Adjust
Adjust