Engström Carestation
6-6
08/07 1505-1018-000
6.3.2 Every 10, 000 hours
In addition to the 12-month requirements, replace the following parts every
10,000 hours. All parts should be replaced before performing the checks,
tests, and calibrations.
#
1. Install 10,000 Hour Kit after 10,000 hours of operation (Check hour
meter).
• P/N 5471002 - 10,000 Hour Kit -230V/50 Hz
• P/N 5471003 - 10,000 Hour Kit - 120 V/60 Hz
10,000 Hour Kit includes:
• Motor OF302
Refer to JunAir Evair 03 TRM - Page 10.
• Non-return valve - 2 each, P/N 5414500
Refer to the JunAir EVair 03 TRM - Figures 4 and 7.
• Solenoid valve
Refer to JunAir EVair03 TRM - Figures 8 and 9.
• Auto drain
Refer to JunAir Evair03 TRM - Page 9.