Engström Carestation
2-2
08/07 1505-1018-000
2.1 Pneumatic Operation
For a complete diagram of the pneumatic system, refer to Figure 11-3, "
Vent
Engine manifold flow diagram
" in Section11.
The EC requires medical-grade oxygen (O
2
) and medical-grade Air sources
ranging from 2.4 to 6.5 bar (35 to 94 psi). Refer to Appendix For additional
information regarding medical-grade gasses. Additional filtering will not be
necessary if supplied gasses are medical grade. If gas supply does not meet
specifications refer to Appendix A for information regarding risks and
recommendations on filters.
The system includes two separate channels (O
2
and Air) to provide dynamic
mixture control of the delivered O
2
percentage.
The Air supply may include an optional compressor unit {1} for applications
where pipeline Air is not available or to provide a continued Air supply if the
pipeline supply goes down.
O2
Air
Air
AB.98.022
{1}
AB.98.055