4 Installation and Service Menus
1505-1018-000 08/07
4-11
4.3 Calibration menu
Note
The
Cal. Flag
menu item is used by the factory to activate the “Calibration
required” alarm. It is set as a reminder that calibrations must be performed
when the machine is set up for operation at its permanent location.
After completing the
O2 FCV
,
Air FCV
, and the
Exhalation Valve
calibrations,
set the
Cal. Flag
to Off.
If EC Service Application is connected, ensure DAC is set to 0 and the
checkbox is marked.
Refer to Section 5.1.1 for further information.
Menu Item
Message text
Comments
O2 FCV
Start O2 Flow Control Valve
calibration and leak test.
Select Start Calibration.
Open inspiratory port to atmosphere.
Start 02 Flow Control Valve calibration.
Air FCV
Start Air Flow Control Valve
calibration and leak test.
Select Start Calibration.
Open inspiratory port to atmosphere.
Start Air Flow Control Valve.
Exhalation Valve
Patient must not be connected to
circuit during calibration. Start
Exhalation Valve calibration.
Connect to the patient circuit.
Occlude the patient wye.
Start Exhalation Valve calibration.
Backlight Test
Start display backlight test.
This test turns off one backlight to test the
other light. Screen brightness may change
during the test.
Gas Calibration
Start gas calibration. Calibrate CO2
and O2 measurements.
Gas Calibration is enabled whenever an
MGAS module is installed. Gas
Calibration is disabled if the MGAS
module is warming up.
Cal. Flag
Turn the Calibration required
message On/Off.
When Cal. Flag is set to On, the
“Calibration required” message is
displayed in the general message area.
Previous Menu
Return to previous menu.
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...