Engström Carestation
7-10
08/07 1505-1018-000
5. Open only the O
2
flow valve by checking the
O2 Flow Valve
box (verify that
60000 DAC counts are set).
6. Repeat the Vent Engine Leak Test.
• If the test fails, check the individual components in the O
2
Flow path.
• If the test passes, check the individual components in the Air Flow path.