Engström Carestation
8-14
08/07 1505-1018-000
8.6.2 VCB input
signal latch
The VCB Input Signal Latch group shows the status of the displayed item.
Item
Description
Module Power
FAIL = Power to the monitor module has failed.
OK = Module Power has not failed or module power has
been manually shut off.
PSM Power
FAIL = Power to the Patient Side Module has failed.
OK = PSM Pwr has not failed or PSM power has been
manually shut off.
Insp Press Zero Valve
ON = Inspiratory pressure zero valve on.
OFF = Inspiratory pressure zero valve off.
Aux Press Purge Valve
ON = Auxiliary pressure purge valve on.
OFF = Auxiliary pressure purge valve off.
Exhalation Vlv Open
YES = VMB is commanding the exhalation valve open.
NO = The valve is not opened.
VMB Buzzer On?
YES = VMB is activating the buzzer.
NO = Buzzer is not active or is activated by the VCB.
Nebulizer Detected?
YES = Nebulizer is connected.
NO = Nebulizer is not connected.
Buzzer Detected?
NO = Buzzer current is too low. Valid when buzzer should
be on.
YES = Buzzer is sensed.
NA = Buzzer not being activated.
Over Pressure
DETECTED = Expiratory over-pressure detected.
NOT DETECTED = Over pressure was not detected.
Vcb Pld Rev
Displays the revsion number of the Programmable Logic
Device software.
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...