Engström Carestation
7-54
08/07 1505-1018-000
VMB Exp Flow Sensor Cal
Pass
Expiratory flow sensor calibration has passed.
None
Waiting for supply pressure
FCV calibration - if a weak supply pressure is
used, system will occasionally pause and wait
for supply pressure to come back up before
resuming FCV cal.
None; normal operation.
Weak Supply Detected1:
Value
FCV calibration - Indicator that system will use
the degraded wall supply routine for FCV
calibration.
None; normal operation. A weak supply
pressure was detected when trying to ramp the
flow up to 165 L/min.
Weak Supply Detected2:
Value
FCV calibration - Indicator that system will use
the degraded wall supply routine for FCV
calibration.
None; normal operation. A weak supply
pressure was detected during the 120 L/min
flow test.
Weak Supply Detected3:
Value
FCV calibration - Indicator that system will use
the degraded wall supply routine for FCV
calibration.
None; normal operation. A weak supply was
detected during final flow ramp-up before first
cal points are taken.
Weak Supply Detected4:
Value
FCV calibration - Indicator that system will use
the degraded wall supply routine for FCV
calibration.
None; normal operation. A weak supply was
detected during final flow ramp-up before first
cal points are taken if DAC output reached
65,000 counts before flow reached 170 L/min.
Xducer Error:
Value
During Paw Xducer check, the insp and exp
pressure difference > 4 cmH
2
O.
Value
= pressure difference.
Refer to Section 5.2.8.
Zero flow samp
Error in FCV calibration cal point projection.
Check for weak or restricted gas pressure supply
Zero for flow sampl2
Error in FCV calibration cal point projection.
Check for weak or restricted gas pressure supply
Zero out of range
ADC counts are out of the acceptable zero
pressure range for the inspiratory and/or
expiratory pressure sensor. The acceptable
range for each sensor is 631 to 968 ADC
counts.
Refer to Section 5.2.8
Zero Pressure Sensors
Airway pressure sensors are currently being
zeroed.
None
Zero Pressure Sensors
Complete
Airway pressure sensor zeros are complete.
None
Message in
“
Calibration running”
field
Reason
Troubleshooting
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...