Engström Carestation
9-2
08/07 1505-1018-000
w
WARNING
To prevent fires:
• Use lubricants approved for O
2
environments, such as Krytox. Other
lubricants may burn or explode in high O
2
concentrations.
• All covers used on the system must be made from antistatic materials. Static
electricity can cause fires in high O
2
concentrations.
w
Obey infection control and safety procedures. Used equipment may contain
blood and body fluids.
w
A movable part or a removable component may present a pinch or a crush
hazard. Use care when moving or replacing system parts and components.
w
Some internal parts have sharp edges and can cause cuts or abrasions. Use care
when servicing internal components.
w
When servicing the EC, extreme care must be taken to avoid introducing foreign
debris into the pneumatic flow passages of the ventilator. This includes dust and
contaminants on the vent engine and particularly metal chips generated by
screw threads. Before removing components on the vent engine, secure a clean
work area and thoroughly clean the vent engine to remove any contaminants.
Failure to do so may result in damage to the internal components.
w
After repairs are completed, always perform the checkout procedure. Refer to
Section 3 of this manual.
w
CAUTION
Electrostatic discharge through circuit boards may damage the components on
them. Wear a static control wrist strap before touching the circuit boards. Handle
all circuit boards by their non-conductive edges. Use anti-static containers when
transporting them.
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...