Engström Carestation
6-8
08/07 1505-1018-000
6.5 Battery capacity test
Although replacement of the backup batteries is recommended at the end of
4 years, batteries that pass the capacity test can be considered viable for
battery backup of the system for up to 6 years at the discretion of the hospital.
Before testing the batteries, ensure that the PMB Diagnostics screen shows
that the battery is fully charged. Battery status will show "Float Charge"
which indicates that the battery is fully charged. Refer to section 8.1.3 .
Test procedure
1. Turn the system on and start a case (simulated).
2. Disconnect the power cord from the mains outlet.
3. Allow the system to run on battery until it does an orderly shutdown.
4. Reconnect the power cord to a mains outlet.
5. Boot the system with the PCMCIA Service Application for the DU and
Compact Flash for HPDU, then access the Power Diagnostics function as
detailed in Section 8.
• On the Main Menu of the Service Application, select
Power Diagnostics
.
• On the Power Diagnostic menu, select
Power Control
.
6. Page 1 of the Power Controller Power Diagnostics screen shows the ‘Date
battery Tested’ (the last full battery discharge until shutdown) and the
‘Discharge Time’.
• If Discharge Time information is not displayed on the screen, the test has
not been run. Repeat the test procedure.
• If the ‘Discharge Time’ is greater than 60 minutes, the batteries can be
left in service for one more year.
• If the ‘Discharge Time’ is less than 60 minutes, both batteries should be
replaced.
N
N
N
Nooo
otttteee
e:::: Charge battery for at least 8 hours before next use.
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...