Engström Carestation
5-12
08/07 1505-1018-000
5.2.6 Vent engine
leak test (high O
2
pressure)
Note: The high O2 pressure leak test will not start if the O2 supply pressure is 35 psig or
below.
1. Verify that the inspiratory outlet is open to atmosphere (not plugged).
2. Open the manual O
2
shut-off valve.
3. On the VMB screen:
• Verify that the measured
O2 Supp
ly
Press
ure = pipeline pressure.
• Verify that the measured
Air Supp
ly
Press
ure = 0.0 psig.
4. On the VCB screen:
• Energize (
!
) the
Effort Valve Energized
valve.
• Energize (
!
) the
Aux
iliary
Press
ure
Purge
valve.
• Verify that the
Exhal
ation
Flow Valve
is energized (
!
).
• Set the
Exh
alation
DAC
counts to 60,000.
• Verify that the
Air
Flow Valve
is energized (
!
).
• Set the
Air DAC
counts to 60,000
(this will release any air pressure in the system).
• De-energize (
"
) the
O2
Flow Valve
.
5. Plug the following ports:
•
INSP
iratory outlet.
•
EXP
iratory inlet.
• Auxiliary pressure port (
Paux
).
6. Close the manual O
2
shut-off valve.
7. On the VCB screen, select
High O2 Leak
to start the test.
8. After the test is complete (Max 15 seconds):
• Verify that the displayed
Leak
rate does not exceed 26 ml/min.
• if not, troubleshoot for leaks in the shaded area.
9. Remove the plugs from the
INSP
iratory,
EXP
iratory, and
Paux
ports.
10.On the VCB screen:
• De-energize (
"
) the
Effort Valve Energized
valve.
• De-energize (
"
) the
Aux
iliary
Press
ure
Purge
valve.
AB.98.086
Содержание Engstrom Carestation
Страница 1: ...Engström Carestation Technical Reference Manual ...
Страница 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Страница 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Страница 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Страница 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Страница 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...
Страница 225: ...ERROR undefined OFFENDING COMMAND STACK ...