background image

USER MANUAL

CAREvent® CA

Part Number: 15PL2151 - REV 11 Feb, 2021

CHEMICAL AGENT ENVIRONMENT AUTOMATIC 

AND MANUALLY TRIGGERED RESUSCITATOR

01CV3000-CA

Содержание CAREvent CA

Страница 1: ...USER MANUAL CAREvent CA Part Number 15PL2151 REV 11 Feb 2021 CHEMICAL AGENT ENVIRONMENT AUTOMATIC AND MANUALLY TRIGGERED RESUSCITATOR 01CV3000 CA...

Страница 2: ...ctional Checks 8 2 1 Set Up 8 2 2 Testing of the Individual Features of the Ventilator 9 3 Operating Procedure 9 3 1 Connecting the Supply Hose Pressure Gauge and Patient Circuit 9 3 2 Manual Ventilat...

Страница 3: ...duals trained in Cardio Pulmonary Resuscitation and the operation of oxygen powered ventilators should use this equipment Thoroughly review this instruction manual before use INDICATIONS FOR USE The C...

Страница 4: ...s warranty gives no specific legal rights You may also have other rights that may vary according to local regulations 1 3 Features The CAREvent CA Handheld Resuscitator is a pneumatically powered time...

Страница 5: ...tentially toxic atmospheres Has a Manually Actuated Automatic Ventilation Override Button Manual Button with a 20 second delay to re start of automatic cycling to allow the provision of 30 chest compr...

Страница 6: ...cmH2 0 2 0 mBar Inspiratory Resistance 6 0 cmH2 0 5 9 mBar Expiratory Resistance 6 0 cmH2 0 5 9 mBar Input Pressure 45 70 PSI 3 1 4 8 Bar Maximum Airway Pressure 60 cmH2 0 58 8 mBar Operating Temperat...

Страница 7: ...cleaned in accordance with the instructions provided in this manual 5 Ensure that all components are reassembled correctly and that all items are replaced in the carrying case 6 After use always ensur...

Страница 8: ...ested during the pre use Functional Check 1 Maximum Airway Pressure and Audible Alarm 2 Frequency Tidal Volume Adjustment 3 Manual Ventilation 4 Demand Valve Function and Automatic Circuit Shut Off 5...

Страница 9: ...nd inflation may increase cardiac output without the danger of gastric distension The flowrate provided is equivalent to the preset automatic flowrate 1 If no respiratory effort is observed position y...

Страница 10: ...the mask indication on pressure gauge will decrease or any obstruction in the patients airway blow off valve will operate reposition patients head and adjust mask and hand position to ensure a good ai...

Страница 11: ...ual button repeatedly or automatically cycling the resuscitator does not clear the foreign material from the patient valve turn the selector to the OFF position remove the facemask and unscrew patient...

Страница 12: ...orking order Regulator working pressure suction if equipped and ventilator limiting pressures should be checked at least every six months and more frequently in high use applications Level I Preventat...

Страница 13: ...clean with a mild soap solution Under no circumstances should the complete unit be allowed to be soaked or immersed in cleaning solutions G Facemask F Pressure Gauge D Manual Button A Resuscitator Bod...

Страница 14: ...ousing from the body of the resuscitator being careful to ensure that the diaphram is retained fig 3 5 Shake out any foreign material 6 Wash all components thoroughly in a mild soap solution and disin...

Страница 15: ...15 CAREvent CA User Manual...

Страница 16: ...In North America call Toll Free 1 800 387 3405 MedNet EC REP GmbH Borkstrasse 10 48163 M nster Germany EC REP SERIAL N0 www otwo com 45A Armthorpe Road Brampton ON Canada L6T 5M4 Telephone 1 905 792 O...

Отзывы: