Performance verification
5-2
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
5.2 Tools, test equipment, and service materials
In addition to the tools listed in Chapter 1, the test equipment and materials listed in
Table 5-2 are required for the performance verification.
Removal/replacement of piston/cylinder assembly,
gear meshing, or greasing of rack
Test 5 (gas volume accuracy, breath timing,
and PEEP system test) and test 6 (EST)
Removal/replacement of motor/encoder assembly,
PEEP pump, or flow sensor assembly
Test 5 (gas volume accuracy, breath timing,
and PEEP system test) and test 6 (EST)
Removal/replacement of oxygen regulator, oxygen
solenoid assembly, oxygen regulator pressure
transducer, mixing manifold, or oxygen sensor
Test 6 (EST) and test 7 (oxygen accuracy
test)
Removal/replacement of safety valve assembly
Test 6 (EST) and test 9 (safety valve test)
Removal/replacement of UI display, pressure solenoid,
controller, or battery backup (BBU) PCB; installation/
removal/replacement of the Communications option
assembly; EPROMs; or NVRAM
Complete performance verification
Removal/replacement of any other part
Test 6 (EST)
Table 5-1: When to run performance verification (continued)
Interval or service performed
Test requirements
Table 5-2: Test equipment required for performance verification
Description
Manufacturer and model or Covidien part
number
Electrical safety analyzer, capable of measuring
ground resistance and leakage current
Fluke ESA620 or equivalent
Pneumatic calibration analyzer or equivalent devices
capable of measuring oxygen percent, flow, BTPS
volume, and pressure. Oxygen analyzer connector
tee. Required accuracies:
Flow: 2.75% of reading ±0.05 sL/min
Volume: 2% of reading or ±1 digit
Low pressure (-150 to +150 cmH
2
O): 0.75% of
reading ±0.04 cmH
2
O
High pressure (0 to 150 psig): 1.0% of reading
±0.1 psi
Oxygen percentage: ±2% oxygen
Puritan Bennett™ PTS 2000 Performance
Test System (4-074686-00)
Digital multimeter (DMM) accurate to three decimal
places
Local supplier
Multimeter patch cord set, 0.025 square receptacle
(Includes red cord
(P/N G-061579-00) and black cord
(P/N G-061580-00))
G-061567-00 or equivalent
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......