Service and repair
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
8-43
Figure 8-38. Exhalation assembly, speaker, and piezo alarm mounting
8.9.4 Exhalation check valve
(See Figure 8-35). Remove the exhalation check valve, as follows. Install by reversing removal
procedure.
1.
Remove exhalation assembly (Section 8.9.3).
2.
Remove heater and thermistor assemblies and spring clip from cross tube to allow access
to screws.
3.
Using 2.5-mm hex key, remove three M3 x 8 screws and split-ring washers from cross tube
(Figure 8-36).
4.
Using 3-mm hex key, remove three M4 x 12 screws and split-ring washers from check
valve housing. Remove gasket, then check valve (Figure 8-39).
NOTE:
• When installing the check valve, make sure the brass knob faces upward (into
ventilator). Installing the check valve backwards prevents the unit from ventilating
properly.
• When reassembling the check valve housing, orient it so the exhalation pressure port
faces at an angle towards the two flow sensor ports.
Underside of ventilator
7-00202
Expiratory
filter
housing
Exhalation
assembly
EXHAUST
port
Serial number
plate
Speaker
Piezo alarm
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......