Self-tests
3-14
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
9
I/E filter test
Checks pressure drop
across inspiratory and
expiratory limbs of
entire patient system.
Checks pressure drop
across filters.
Fault if not
passed
Yes
Mandatory in EST, but optional
in SST.
Disconnect and reconnect
inspiration (I) tube from/to
inspiratory filter outlet, and
exhalation (E) tube from/to
expiratory filter inlet, as
directed. Block and unblock wye
with a no. 2 stopper, as directed.
Press ACCEPT to signify you have
performed each requested
action.
NOTE:
Overriding this fault could result in inadequate bacteria protection or excessive resistance to inspiration
or exhalation.
10
PEEP system test
Verifies PEEP system can
generate and maintain
preset PEEP levels
Failure if not
passed
Yes, but
not
identical
Verifies PEEP system can
generate and maintain preset
PEEP levels within either of two
sets of limits. If results are within
the inner set of limits, test passes
the first time. If the results are
outside of the outer set of limits,
test fails the first time. If the
results are outside the inner
limits but within the outer limits,
the ventilator adjusts its
calibration table in nonvolatile
RAM (NVRAM) and repeats the
test to verify the calibration. The
test fails if the calibration cannot
be verified after five attempts.
11
Check valve test
Checks check valve
outlet port of cylinder
Failure if not
passed
No
12
Piston leak test
Checks piston leak
against a table of leak
values stored in NVRAM
Failure if not
passed
No
First remove inspiratory filter;
then block and unblock
inspiratory filter port as
directed, using no. 3 stopper.
Press ACCEPT to signify you have
performed each requested
action. Replace inspiratory filter
at end of test.
13
Lights-displays
test
Verifies lights and
displays function
Failure if not
passed
No
Observe each named display,
and press ACCEPT to verify it is
lit. Press CLEAR if a display is not
lit.
14
Keys test
Verifies every row and
column of key matrix
can be correctly read
Failure if not
passed
No
Press each named key (each key’s
LED will also be lit). Press CLEAR
if a key’s LED is not lit.
Table 3-8: EST tests (continued)
No.
Test
Purpose
Status
In SST?
Comments
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......