Diagnostic codes
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
6-25
9013
Ongoing hardware
checks
Main fan failure
1. Make sure room temperature is not too
low. Make sure that ventilator has
warmed up and attained a steady-state
temperature.
2. Make sure ventilator lid is closed.
3. Make sure vents in unit are not
blocked.
4. Clean or replace fan filter, as necessary.
5. Check that main fan/ventilator head
harness connection is secure.
6. Replace main fan.
7. Replace air flow thermistor assembly.
9014
Ongoing hardware
checks
A VENT INOP condition was
declared
1. See previous diagnostic codes in log for
corrective actions.
2. Run EST.
3. Replace controller PCB.
9015
Ongoing hardware
checks
Not Ventilating signal used in
UI/BD communications is not in
expected state
1. See Section 6.2.
2. Replace controller PCB.
9016
Ongoing hardware
checks
Main alarm speaker failure
1. Check wiring to speaker.
2. Replace speaker.
3. Replace pressure solenoid PCB.
9017
Ongoing hardware
checks
Safety valve current problem
1. Replace safety valve.
2. Replace pressure solenoid PCB.
9018
Ongoing hardware
checks
Safety valve current problem
1. Replace safety valve.
2. Replace pressure solenoid PCB.
9019
Ongoing hardware
checks
Safety valve current problem
1. Replace safety valve.
2. Replace pressure solenoid PCB.
9020
Ongoing hardware
checks
Safety valve current problem
1. Replace safety valve.
2. Replace pressure solenoid PCB.
9021
Ongoing hardware
checks
Safety valve failure
1. Replace safety valve.
2. Replace pressure solenoid PCB.
9022
Ongoing hardware
checks
Safety valve failure
1. Replace safety valve.
2. Replace pressure solenoid PCB.
9023
Ongoing hardware
checks
ALARM light failure
1. Replace UI display PCB.
2. Replace controller PCB.
9024
Ongoing hardware
checks
ALARM light failure
1. Replace UI display PCB.
2. Replace controller PCB.
9025
Ongoing hardware
checks
Pressure solenoid PCB serial
EPROM failure
Replace pressure solenoid PCB.
9026
Ongoing hardware
checks
Pressure solenoid PCB serial
EPROM failure
Replace pressure solenoid PCB.
9027
Ongoing hardware
checks
Pressure solenoid PCB serial
EPROM failure
Replace pressure solenoid PCB.
9028
Ongoing hardware
checks
Pressure solenoid PCB serial
EPROM failure
Replace pressure solenoid PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......