Self-tests
3-10
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
NOTE:
It may be useful to complete EST even with errors, because information on multiple
errors can facilitate troubleshooting.
Table 3-5: Key functions during EST
Key
Function
ACCEPT
Ignore failure and continue.
CLEAR
Repeat a test or return to prompt at start of a test.
Alarm reset
Retest from beginning of EST (when all tests are completed).
Alarm silence
Stop testing and skip to end of EST.
NOTE:
The Alarm silence key is disabled under these conditions: (1) when EST is run as a result of a VENT INOP
condition, and (2) (temporarily until the Circuit comp test is complete) if the tubing type or humidifier type
you’ve selected has changed from the previous time EST was run.
Manual insp
Use to override (when all tests are completed). Can only be used if EST has faulted
or testing is incomplete.
Table 3-6: Prompts during EST testing
Prompt
Operator action
Disconnect O
2
supply
Disconnect ventilator from oxygen supply. Press ACCEPT.
Reconnect O
2
supply
Reconnect ventilator to oxygen supply. Press ACCEPT.
Disconnect I tubing
Disconnect tubing from inspiratory filter outlet. Leave inspiratory filter in place.
Press ACCEPT.
I drop =
x x.x
at 100
or
I drop =
x x.x
at 40
Reconnect I tubing
This is the inspiratory filter pressure drop in cmH
2
O at a flow of 100 L/min (adult
tubing) or 40 L/min (pediatric tubing).
Reconnect tubing to inspiratory filter outlet.
Press ACCEPT.
Unblock wye
Remove stopper from patient wye. Press ACCEPT.
Block wye
Insert a stopper into patient wye. Press ACCEPT.
Disconnect E tubing
Disconnect tubing from expiratory filter inlet. Leave expiratory filter in place. Press
ACCEPT.
E drop =
x.x
at 100
or
E drop =
x
x.x
at 40
Reconnect E tubing
This is the expiratory filter pressure drop in cmH
2
O at a flow of 100 L/min (adult
tubing) or 40 L/min (pediatric tubing). Reconnect tubing to expiratory filter outlet.
Press ACCEPT.
Block insp filt port
Remove inspiratory filter. Insert no. 3 stopper into inspiratory filter outlet. Press
ACCEPT.
Reconnect I tubing
Remove stopper from inspiratory filter outlet. Reinstall inspiratory filter. Press
ACCEPT.
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......