Diagnostic codes
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
6-35
9170
Ongoing hardware
checks
Piston did not move forward
for three consecutive breaths
1. Run SST or EST to make sure ventilator
is capable of moving piston.
2. Check harness from controller PCB to
motor/encoder.
3. Replace controller PCB.
4. Replace motor/encoder assembly.
9171
Ongoing hardware
checks
EPROM on pressure solenoid
PCB uninitialized or data
corrupted
Replace pressure solenoid PCB.
9172
Ongoing hardware
checks
EPROM on pressure solenoid
PCB uninitialized or data
corrupted
Replace pressure solenoid PCB.
9173
Ongoing hardware
checks
NVRAM failure
1. Replace NVRAM.
2. Replace controller PCB.
9174
Ongoing hardware
checks
NVRAM failure
1. Replace NVRAM.
2. Replace controller PCB.
9175
Ongoing hardware
checks
NVRAM failure
1. Replace NVRAM.
2. Replace controller PCB.
9176
Ongoing hardware
checks
NVRAM failure
1. Replace NVRAM.
2. Replace controller PCB.
9177-9180
POST
NVRAM failure
1. Replace NVRAM.
2. Replace controller PCB.
9181
Ongoing hardware
checks
The voltage used to identify
the 740/760 user interface is
out of range.
1. Check all cable connections between
the user interface and controller PCB.
2. Replace the user interface.
3. Replace the controller PCB.
9182-9185
POST
NVRAM failure
1. Replace NVRAM.
2. Replace controller PCB.
9186
Ongoing hardware
checks
Bad DAC value during P
e
autozero
Replace controller PCB.
9187
Ongoing hardware
checks
Bad DAC value during P
cyl
autozero
Replace controller PCB.
10000
Interrupt routines
Software error
10001
Interrupt routines
Software error
10002
Interrupt routines
1. Microcontroller failure
2. Software error
1. Replace controller PCB.
2. See Section 6.2.
10003
Interrupt routines
Memory or microcontroller
failure
Replace controller PCB.
10004
Interrupt routines
Microcontroller failure
Replace controller PCB.
10005
Interrupt routines
Microcontroller failure
Replace controller PCB.
10006
Interrupt routines
1. Software error
2. Microcontroller failure
1. See Section 6.2.
2. Replace controller PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......