Service and repair
8-42
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
8.9.2.2 Installing exhalation heater and thermistor assemblies
1.
(See Figure 8-35 and Figure 8-36). Install heater assembly by reversing removal
procedure, wrapping long heater around top of flow sensor housing and short heater
around cross tube.
2.
Install thermistor assembly:
a.
Install black thermistor bead by threading bead and approximately 15 mm of
thermistor wire through hole in clip from outside (Figure 8-37). Then, install clip on
flow sensor housing.
b.
When installing thermistor housed within metal tab, attach tab to inner flow sensor
screw (near flat edge of flow sensor assembly). Angle thermistor tab to obtain
maximum contact with exhalation valve metal housing.
Figure 8-37. Threading thermistor through flow sensor heater clip
8.9.3 Exhalation assembly
(See Figure 8-35). Remove the exhalation assembly as follows. Install by reversing removal
procedure.
1.
Remove exhalation assembly cover (Section 8.9.1).
2.
Remove expiratory filter.
3.
Disconnect heater and thermistor harnesses from ventilator head harness.
4.
Remove all tubes from exhalation assembly (two tubes from both sides of flow sensor,
one tube from check valve, and one tube from exhalation valve).
5.
Remove four M3 x 8 POZIDRIV™* screws with captive washers from around gas exhaust
(Figure 8-38). Remove three M3 x 8 POZIDRIV™* screws with captive washers, and nuts
with captive lockwashers from expiratory filter housing. Remove exhalation assembly.
NOTE:
When reinstalling exhalation assembly, take care to install all tubes in correct positions.
7-00201
Thermistor
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......