Diagnostic codes
6-6
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
6008
BIOT (technical alert --
KEYBOARD FAILED)
A key was held down longer
than expected
1. Verify keyboard cable is securely
connected to UI display PCB.
2. Replace keyboard.
3. Replace controller PCB.
4. Replace UI display PCB.
NOTE:
Holding down any key other than MENU during POST causes error code
6008.
6009
BIOT (technical alert --
LOW SYS TEMP
ALERT)
Temperature inside ventilator
enclosure too low
1. Make sure room temperature is not too
low. Make sure that ventilator has
warmed up and T
box
attained a steady
state.
2. Replace pressure solenoid PCB.
6010
BIOT (technical alert --
HI SYS TEMP ALERT)
Temperature inside ventilator
enclosure too high
1. Make sure room temperature is not too
high.
2. Make sure vents in unit are not
blocked.
3. Make sure main fan is operating
properly. Clean or replace fan filter, as
necessary.
4. Replace pressure solenoid PCB.
6011
BIOT (technical alert --
EXH CCT LOW TEMP)
Exhalation limb temperature
too low
1. Check connections of exhalation heater
and thermistor assemblies.
2. Replace exhalation heater assembly.
3. Replace exhalation thermistor
assembly.
4. Replace pressure solenoid PCB.
NOTE:
Be aware that spirometry may be affected (monitored values may be
higher than actual).
6012
BIOT (technical alert --
EXH CCT HI TEMP)
Exhalation limb temperature
too high
1. Check connections of exhalation heater
and thermistor assemblies.
2. Replace exhalation heater assembly.
3. Replace exhalation thermistor
assembly.
4. Replace pressure solenoid PCB.
NOTE:
Be aware that spirometry may be affected (monitored values may be
higher than actual).
6013
BIOT (technical alert --
FLO SENSOR LO TEMP)
Temperature at exhalation
flow sensor pressure
transducer too low
1. Make sure room temperature is not too
low. Make sure that ventilator has
warmed up and T
xdcr
attained a steady
state.
2. Replace pressure solenoid PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......