Service and repair
8-18
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
8.5.5 LCD panel (message window)
Caution
When disassembling the UI, account for all fasteners removed, as fasteners falling
between the keyboard and the UI display PCB may cause equipment failure.
Remove the LCD panel as follows. Install by reversing removal procedure.
1.
2.
Using 5-mm nutdriver, remove four M2.5 nuts, internal lockwashers, and fiber washers
that attach LCD panel to UI display PCB. Remove LCD panel (Figure 8-13).
3.
Detach LCD panel ribbon cable from UI display PCB (Figure 8-12).
NOTE:
When installing the LCD panel, be sure ribbon cable is at top of panel.
Figure 8-13. Removing LCD panel
7-00187
LCD panel
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......