Diagnostic codes
6-30
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
9096
Ongoing hardware
checks
Piston motion error
1. Verify operation of the home
optoswitch.
2. Replace controller PCB.
3. Replace motor/encoder assembly.
9097
Ongoing hardware
checks
Piston motion error
1. Verify operation of the end optoswitch.
2. Replace controller PCB.
3. Replace motor/encoder assembly.
9098
Ongoing hardware
checks
Motor controller chipset
failure
Replace controller PCB.
9099
Ongoing hardware
checks
Piston motion error
Replace controller PCB.
9100
Ongoing hardware
checks
Piston motion error
Replace controller PCB.
9101
Ongoing hardware
checks
Piston motion error
1. Check gear pin. If piston is loose,
replace piston/cylinder assembly.
2. Replace BBU PCB.
9102
Ongoing hardware
checks
Piston motion error
1. Replace piston/cylinder assembly.
2. Replace BBU PCB.
9103
Ongoing hardware
checks
Piston motion error
1. Check that home optoswitch/ventilator
head harness connection is secure.
2. Replace controller PCB.
9104
Ongoing hardware
checks
Piston motion error
1. Check that end optoswitch/ventilator
head harness connection is secure.
2. Replace controller PCB.
9105
Ongoing hardware
checks
Piston motion error
Replace controller PCB.
9106
Ongoing hardware
checks
Atmospheric pressure out of
bounds
1. Make sure ventilator is operating
within stated environmental
specifications.
2. Replace pressure solenoid PCB.
9107
Ongoing hardware
checks
Motor controller chipset
failure
Replace controller PCB.
9108
Ongoing hardware
checks
Not Ventilating signal used in
UI/BD communications is not in
expected state.
1. See Section 6.2.
2. Replace controller PCB.
9109
Ongoing hardware
checks
Not Ventilating signal used in
UI/BD communications is not in
expected state.
1. See Section 6.2.
2. Replace controller PCB.
9110
Ongoing hardware
checks
Exhalation solenoid seems to
be stuck open
1. Replace exhalation solenoid.
2. Replace pressure solenoid PCB.
9111
Ongoing hardware
checks
CAUTION light control bit not
in expected state
1. Replace UI display PCB.
2. Replace controller PCB.
9112
Ongoing hardware
checks
CAUTION light control bit not
in expected state
1. Replace UI display PCB.
2. Replace controller PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......