Performance verification
5-12
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
7.
Change the ventilator setting shown in
bold
.
8.
Make the following measurements, and verify the readings are as listed below.
See
Section 5.7 15, 16, 17
.
5.6.7 Extended self-test (EST) (Test 6)
Warning
• Before running EST, you must disconnect the ventilator from the patient. Running
EST while the ventilator is connected to the patient can injure the patient.
• Do not use or override a ventilator that completes EST with a fault status without
first verifying its operational readiness by an independent means and determining
that a patient will not be placed at risk.
• Covidien urges medical departments to review the implications of using a ventilator
that completed EST with a fault status. Covidien recommends establishing a medical
department protocol that defines the conditions under which ventilator usage is
acceptable.
Control
Setting
HIGH RATE
100 breaths/min
HIGH TIDAL VOLUME
1500 mL
LOW INSP PRESSURE
3 cmH
2
O
LOW TIDAL VOLUME
0 mL
HIGH PRESSURE
90 cmH
2
O
LOW MINUTE VOLUME
0 L
Mode
A/C, VCV (SQUARE flow waveform for software
revision J or later)
RESPIRATORY RATE
10 breaths/min
TIDAL VOLUME
1000 mL
PEAK FLOW
80 L/m
PLATEAU 0
s
PEEP/CPAP
20 cmH
2
O
TRIGGER SENSITIVITY
20 L/min
% O
2
21 %
Measure the following
Acceptable range
Tidal volume
890 to 1110 mL
Respiratory rate
9 to 11 breaths/min
PEEP
17.39 to 22.61 cmH
2
O
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......