Diagnostic codes
6-34
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
9162
Ongoing hardware
checks
Oxygen regulator pressure
transducer reading too low
1. Calibrate oxygen regulator pressure
transducer.
2. Replace oxygen regulator pressure
transducer.
3. Replace oxygen regulator assembly.
4. Replace pressure solenoid PCB.
9163
Ongoing hardware
checks
Oxygen regulator pressure
transducer reading too high
1. Verify oxygen supply pressure is not too
high.
2. Calibrate oxygen regulator pressure
transducer.
3. Replace oxygen regulator pressure
transducer.
4. Replace oxygen regulator assembly.
5. Replace pressure solenoid PCB.
9164
Ongoing hardware
checks
Safety valve stuck closed
1. Check that safety valve/ventilator head
harness connection is secure.
2. Replace safety valve.
3. Replace pressure solenoid PCB.
9165
Ongoing hardware
checks
Safety valve stuck closed
1. Check that safety valve/ventilator head
harness connection is secure.
2. Replace safety valve.
3. Replace pressure solenoid PCB.
9166
Ongoing hardware
checks
Atmospheric pressure
measurement out of range
1. Make sure ventilator is operating
within stated environmental
specifications.
2. Verify tube between inspiration
manifold and inspiration pressure
transducer is securely connected.
3. Replace pressure solenoid PCB.
4. Replace controller PCB.
9167
Ongoing hardware
checks
Atmospheric pressure
measurement out of range
1. Make sure ventilator is operating
within stated environmental
specifications.
2. Verify tube between inspiration
manifold and inspiration pressure
transducer is securely connected.
3. Replace pressure solenoid PCB.
4. Replace controller PCB.
9168
Ongoing hardware
checks
Motor controller chipset
failure
Replace controller PCB.
9169
Ongoing hardware
checks
Piston did not move forward
for three consecutive breaths
Ventilator has Rev J software
1. Make sure that Rev K or later software
is installed in ventilator.
2. Run SST or EST to make sure ventilator
is capable of moving piston.
3. Check harness from controller PCB to
motor/encoder.
4. Replace controller PCB.
5. Replace motor/encoder assembly.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......