Theory of operation
2-54
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
Calibration and other
constants
Number of motor quadcounts between
home and end positions
Must be input when new piston/cylinder
installed.
Calibration and other
constants
Oxygen mixing constants, including flows
from orifices and regulator pressures
Must be input when new oxygen
regulator assembly installed.
Calibration and other
constants
PEEP pump calibration lookup table
Source: PEEP pump calibration. Lookup
table updated during EST and calibration
verified.
Calibration and other
constants
Current of PEEP pump at 0% duty cycle
Value in counts. Currently unused.
Calibration and other
constants
Oxygen regulator pressure transducer
calibration data
Source: oxygen regulator pressure
transducer calibration. Checked during
EST.
Calibration and other
constants
Oxygen sensor (FIO
2
) calibration data
Calibration verified during FIO
2
calibration check.
Calibration and other
constants
Pressure solenoid calibration constraints
Updated when new pressure solenoid
board is installed and during calibration.
Operator accesses via menu.
Ventilator configuration
Ventilator breathing circuit resistance
and compliance
Source: SST/EST (circuit resistance field
currently unused).
Ventilator configuration
Endotracheal tube size, in mm
Source: SST/EST
Ventilator configuration
Ventilator breathing circuit type
Source: SST/EST
Ventilator configuration
Model
Ventilator configuration
Humidification device type
Source: SST/EST
Ventilator configuration
Circuit characteristic change
Indicates whether there has been a
change to the circuit type, humidification
device, or ET tube size, used to determine
whether circuit compliance test must be
run before skipping to the end of SST or
EST.
Ventilator options
Nebulizer settings
Most recently selected nebulization
phase and length of nebulization time
Ventilator options
Serial port configuration
Most recently selected settings for serial
ports A and B
Operator settings
Breath settings
Operator settings
Alarm settings
Operator settings
Apnea settings
Operator settings
Main alarm volume
Range: 1 - 5
Operator settings
Pressure units
Range: cmH
2
O or hPa
Operator settings
Date format selection
Range: Month or day first (US or
European)
Operator settings
Time format selection
Range: 12- or 24-hour clock
Table 2-2: NVRAM contents (continued)
Type of data
Parameter
Notes
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......