Service and repair
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
8-39
NOTE:
• If the monitored oxygen concentration of delivered gas is too low, the outlet or inlet
check valve may be stuck open. Apply negative pressure to the wye. If system
pressure stays low, check for a stuck check valve.
• When installing the outlet check valve, make sure the leaf faces toward outside of
ventilator. Installing the check valve backwards prevents the unit from ventilating.
• To prevent the check valve and gasket from falling out during installation, try
inclining the ventilator slightly.
Figure 8-33. Removing inspiration manifold assembly
8.9 Exhalation assembly
8.9.1 Exhalation assembly cover
(See Figure 8-34). Remove the exhalation assembly cover as follows. Install by reversing
removal procedure.
1.
Open and prop lid (Section 8.5.1).
2.
Remove options panel plate (Section 8.6.2).
3.
Remove filter from reservoir assembly.
4.
Disconnect pilot tube from side of exhalation valve.
5.
Using 4-mm hex driver, remove M5 x 12 screw and flat washer that retain cover and
retaining clip. Remove clip. Lift off cover.
7-00241
Gasket
Inspiration
manifold
assembly
Outlet check valve
(leaf toward outside)
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......