Service and utility functions
4-12
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
If you selected
Exit
testing, do the following:
Figure 4-2. Oxygen regulator pressure transducer calibration setup
O
2
pressure calib
Verify reg pressure
xx.x
psi
x.xxx
bar
Press ACCEPT if the value displayed matches the pressure measurement device
reading entered above, ±0.2 psi; otherwise, press CLEAR.
O
2
pressure calib
Disconnect test gauge
Remove pressure measurement device and
replace cap
. Press ACCEPT.
O
2
pressure calib
Calibration passed
or
Calibration failed
If the calibration passes, press ACCEPT to return to the
Calibration tests
menu. If
you do not intend to perform any other calibration functions, turn the knob to
access
Exit testing
; otherwise, turn the knob to access another calibration function.
If the calibration fails, try the following corrective actions:
1. Verify oxygen supply pressure is between 40 and 90 psi (275 and 620 kPa).
2. Repeat this test and verify that oxygen is disconnected and reconnected when
prompted.
3. Verify oxygen regulator pressure transducer is securely connected to main
ventilator head harness.
4. Replace oxygen regulator pressure transducer and recalibrate.
5. Replace pressure solenoid PCB.
NOTE:
If the ventilator was functioning correctly before service, but the calibration
fails, it is unlikely that a PCB is malfunctioning.
Message window display
Description
Message window display
Description
Exit testing
Press ACCEPT to reset the ventilator and exit the service menu in preparation for
ventilation.
7-00171
PTS 2000 or other pressure
measurement device (set
to read pressure)
Oxygen regulator
pressure sensor
calibration tool
High Pressure
Range P+ port
NOTE: Remove cap from oxygen
regulator pressure adapter.
Replace cap
when finished.
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......