Theory of operation
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
2-33
Feedback from the exhalation pressure transducer helps the ventilator maintain the PEEP/
CPAP at the selected value. Even with a ventilator breathing circuit leak of up to 10 L/min,
PEEP/CPAP can be maintained. The ventilator compensates by adjusting the piston/cylinder
output.
Figure 2-20. PEEP/CPAP system
PEEP/CPAP system
Filter
7-00111
PEEP pump
PEEP reservoir
Tubes crossed
Location of PEEP/CPAP system
PEEP/CPAP system
7-00112