Service and repair
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
8-57
8.12 Piston/cylinder and motor/encoder assemblies
Caution
Never attempt to open up the piston/cylinder assembly. Opening up the piston/cylinder
may allow debris into the cylinder, causing damage.
8.12.1 Optoswitches
(See Figure 8-45). Remove the optoswitches as follows. Install by reversing removal
procedure.
1.
Remove exhalation assembly (Section 8.9.2).
2.
Remove options panel plate (Section 8.6.2).
3.
Disconnect main ventilator head harness from optoswitch.
4.
Using 1.5-mm hex driver, remove four M3 x 8 screws and split-ring washers per
optoswitch. Remove optoswitch.
Figure 8-45. Piston/cylinder assembly installed
7-00035
Optoswitch
assembly
(2 PL)
Rack cover
retaining screws
and washers
Piston
cylinder
assembly
Captive
knurled
screw (2 PL)
Motor/encoder
Содержание Puritan Bennett 700 Series
Страница 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Страница 18: ...xviii 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 24: ...xxiv 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 160: ...3 22 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 186: ...4 26 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 398: ...9 64 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 414: ...A 16 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 438: ...12 10070389 Rev A Puritan Bennett 700 Series Ventilator System Service Manual This page intentionally blank...
Страница 439: ......