Operational Verification
6-6
Service Manual
6.5.2
Circuit Leak Test
Figure 6-3.
Circuit Leak Test
Note:
The circuit check test can only be performed in ventilation standby.
1.
Remove test lung, if present, and occlude the end of the patient circuit Y-piece using CAP100P.
2.
Turn on the ventilator.
3.
Touch the
Circuit Check
button on the screen, and follow the on-screen instructions.
4.
After the test is completed, verify that the message “PASSED” is displayed. Document the results on the
record sheet.
5.
Reattach the test lung.
6.
To exit ventilation standby, touch the
Start Ventilating
button on the screen.
6.5.3
Exhalation Flow Sensor and FiO
2
Sensor Calibration
1.
Touch the
Sensors
button on the screen.
2.
Touch the
Flow Sensor
button on the screen, and follow the on-screen instructions.
Содержание Newport e360
Страница 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Страница 7: ...v C Diagrams Index...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank viii...
Страница 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Страница 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Страница 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Страница 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Страница 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Страница 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Страница 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Страница 152: ...Page Left Intentionally Blank 1 2...
Страница 153: ......